Barcelona – ITMA 2011

Newton at ITMA 2011
Newton had an excellent trip to Barcelona, Spain for the ITMA 2011 textile exhibition. As a booth companion, I can say he was definitely the strong, silent type – not much for conversation but quite popular nonetheless. Sharing the spotlight with Newton were Measurement Technology NW’s SGHP-8.2 sweating hotplate (skin model), and our new RPP (Radiant Protective Performance) device for assessing a fabric’s resistance to radiant heat energy.
All these items garnered significant interest and proved that ITMA is the premier show for the textile industry. To call it a huge exhibition is an understatement. It was a full 8 days long, with attendance totaling well over 100,000 people. Every facet of the textile industry is represented: from raw materials, to finished goods, to everything in-between.
Next up for MTNW is the IFAI Expo in Baltimore (October 25-27), where Newton will put on a repeat performance. The SGHP-8.2 unit we had on display at ITMA remained in Europe, but at IFAI it will be replaced with our TPP (Thermal Protective Performance) device.
Interested? Come on by! You’ll find Measurement Technology NW in Booth 2509.
Tags: exhibition, ITMA, ITMA 2011, measurement technology northwest, MTNW, Newton, SGHP, SGHP hotplate, sweating manikin, textile, thermal manikin
This entry was posted
on Tuesday, October 11th, 2011 at 11:33 am and is filed under Thermal Measurement.
You can follow any responses to this entry through the RSS 2.0 feed.
Responses are currently closed, but you can trackback from your own site.
New Year’s Resolutions
With the beginning of a new year comes New Year’s resolutions, and for all our sweating manikin operators we heartily recommend making the resolution to pay more attention to your manikin’s fluid circulation system. Towards that goal I have re-posted a service bulletin from a few years ago that describes the maintenance procedures our engineers have developed for the proper care of our sweating manikins – including Newton (all models) and the Thermal Hand, Thermal Foot, and Thermal Head systems. Note: The procedures shown below apply only to MTNW manikins using removable wicking fabric skins. For manikins with MTNW’s porous metal sweating skin surfaces, please contact MTNW for maintenance information specific to these systems.

MTNW Service Bulletin
Sweating Manikin Fluid System Storage and Flushing Procedures
Long-term storage of water systems, even with high purity water, can result in bacterial growth and a resulting biofilm on all wetted surfaces. Studies have demonstrated significant bacterial activity within 7 days in a high-purity closed system. The large number of variables affecting this phenomenon, including supply water purity, existing bacterial content of water system, ambient temperature, and wetted materials, can make each location unique.
MTNW’s sweating manikin systems meet many of the criteria for promoting bacterial growth. The sealed design of MTNW’s fluid system does not permit reliable purging of the internal tubes, so the system remains wetted throughout its service life. The flowrates are low and often stagnant during non-sweating operation. Water recirculates through the system, so any bacterial content developed in the manikin tubes gets returned to the reservoir. Adding more high-purity water to the reservoir does not eliminate any existing biological activity.
The maintenance procedures outlined below are intended to minimize/reduce bacterial growth, and also offer a shock treatment to systems for periodic cleaning. NOTE: All supply water should be deionized or distilled high purity water.
If you will be USING the manikin sweating system AT LEAST ONCE PER WEEK:
- Fully drain and rinse out the manikin fluid reservoir with supply water
- Refill reservoir with supply water
- Unplug recirculation (black) tube from reservoir and route it to a drain or collection bucket
- Turn on the fluid pump and allow it to run for 10 minutes to purge the old water from the system
- Top off the reservoir and resume testing
If you will be STORING the manikin or not operating the fluid system (re: non-sweating tests) for MORE THAN ONE WEEK:
- Remove the wicking fabric sweating suit
- Add 5 liters of supply water to empty reservoir
- Add 1 ml of unscented household bleach and shake the reservoir to mix and coat the inside surfaces (this will result in approximately 10 PPM sodium hypochlorite solution)
- With the manikin tubing fully connected, turn on the fluid pump and allow it to run for 10 minutes, recirculating this solution through the closed system.
- Set all the manikin sweat rates to 2000 ml/hr/m2, and allow the system to sweat for 10 minutes.
- The system can now be stored for up to 2 months containing this solution
- Prior to running sweating tests, repeat this procedure with fresh supply water to purge chlorine from the system
Every TWO MONTHS as preventative maintenance or as a special one-time shock treatment after storage:
- Very important – Long-term exposure to sodium hypochlorite can degrade internal manikin system components. Do not allow bleach solution to remain inside the manikin system for more then 10 minutes.
- Caution – This process uses bleach with a concentration which could possibly discolor fabrics or cause staining. Use colorfast or white absorbent pad and towels for this procedure. Wear eye protection.
- Remove the wicking fabric sweating suit and place an absorbent pad beneath manikin to catch runoff
- Add 5 liters of supply water to empty reservoir
- Add 100 ml of unscented household bleach and shake the reservoir to mix and coat the inside surfaces (this will result in approximately 1000 PPM sodium hypochlorite solution)
- With the manikin tubing fully connected, turn on the fluid pump and allow it to run for 5 minutes, recirculating this solution through the closed system.
- Set all the manikin sweat rates to 2000 ml/hr/m2, and allow the system to sweat for 5 minutes.
- Set sweat rates back to 0 ml/hr/m2 and quickly blot the manikin dry with a colorfast cloth and turn off fluid pump
- Immediately drain and rinse reservoir, and refill with clean supply water
- Unplug recirculation (black) tube from reservoir and route it to a drain or collection bucket
- Turn on the fluid pump and allow it to run for 5 minutes to purge the system
- Set all the manikin sweat rates to 2000 ml/hr/m2, and allow the system to sweat for 5 minutes.
- Set sweat rates back to 0 and blot manikin dry with colorfast cloth
- Allow the pump to run for an additional 10 minutes, flushing water out the recirculation (black) tube.
- Begin using the manikin for sweating tests within one week, or follow the previously mentioned storage procedure